In this research a one dimensional position sensitive gas detector is designed and constructed for x-ray detection. The inlet window is linear with 14 centimeter length, 5 millimeter height and depth of active area is 8 millimeter. A thin metal blade used as anode and copper strips is used as segmented cathode.
When an x-ray photon becomes incident on the detector anode, induce electrical charges at a position that is spatially coincident with the incident photon. The position of the x-ray is determined by the center of gravity method. This detector is used in proportional mode with a gas mixture of argon (85%) and ethane (15%).
The experiments for position sensivity is tested with x-ray tube (220 watts of power) in various pressures and results is indicated good spatial resolution can be achieved.
Keywords: gaseous detector, metal blade, copper strips, proportional mode, center of gravity, x-ray, spatial resolution
چکیده (به فارسی)
در این پژوهش، یک آشکارساز گازي حساس به مکان یک بعدي طراحی و ساخته شد. پنجره ورودي
آشکارساز به صورت خطی با طول 14 سانتی متر، ارتفاع 5 میلی متر و عمق ناحیه فعال آن 8 میلی متر می باشد. از یک تیغه فلزي نازك به عنوان آند و از نوارهاي مسی به عنوان هر بخش مجزاء کاتد استفاده گردید.
هنگامی که یک فوتون ایکس در مقابل آند فرود می آید، بار الکتریکی در موقعیتی که متناظر با موقعیت فوتون ورودي می باشد، بر روي نوارهاي کاتدي القا شده و موقعیت پرتو ایکس با استفاده از روش مرکز جرم محاسبه می گردد.آشکارساز ساخته شده در مد تناسبی با مخلوط گازي انتخابی آرگون 85 درصد و اتان 15 درصد، مورد استفاده قرار گرفت.
آزمایش هاي حساس به موقعیت با استفاده از منبع پرتو ایکس با آند تنگستنی و با توان 220 وات، انجام گردید و نتایج حاکی از آن است که به تفکیک فضایی مناسب می توان دست یافت.
کلمات کلیدي :آشکارساز گازي، تیغه ی فلزی، نوارهاي مسی، مد تناسبی، مرکز جرم، پرتو ایکس، تفکیک فضایی